Pressure vessel inspection
Renewal interval
Recommended high pressure hose renewal interval:
7 years
Visual inspection
Pressure vessel visual inspection interval:
5 years
Function test inspection
Pressure vessel pressure test interval:
10 years
* In Germany those intervals are obligatory by law. Please do check your local regulations to verify the required inspection and pressure testing intervals of your country.
Calibration intervals
Calibration should be carried out by your local or national calibration service. It is left to your quality management system to define, how often to calibrate gauges, transducers, and electronics. However, here is what we would recommend:
Gauges
Gauges of class 0.6 and 1.0:
1 - 2 years
Pressure transducers
1 - 2 years
Get a quote from your calibration service first to evaluate, if a new gauge or transducer with certificate might be cheaper than the calibration service. If you are outside Europe, it will be our pleasure to handle calibration with the OEM on your behalf.
Force sensors
As long as the zero point of the force sensor is within 3% off the range, the sensor can be considered as good. Check the zero point frequently to be sure about your force sensor.
METRUS service technician
Many METRUS clients have their test bench maintained frequently by a METRUS service technician. Continue to contact page